| Mechanism: |
Linear peristaltic volumetric infusion pump |
| Dimensions: |
5.5"W x 9.4"H x 7.8"D |
| Weight: |
6.8 lbs |
| Power Requirements: |
100-120 VAC, 50-60 Hz, 13 Watts |
| Leakage Current: |
Max 30uA |
| Battery: |
7.2V, 1.2 Ah NiCd (rechargeable) |
| Battery Operating Time: |
3.5 hours at 125 mL/hr |
| Rate Range: |
0.1 to 800 mL/hr in 0.1 mL |
| Volume to be Infused Range: |
0.1 to 9999.9 mL |
| Occlusion Pressure: |
Variable Settings Low: 8 PSI High: 17 PSI |
| KVO Rate: |
1 mL/hr or selected rate, whichever is lower |
| Flow Rate Accuracy: |
± 5% measurement time 8hr, temp. 71.6º F |
| Alarms: |
Air-in-Line, Low Battery, Plug In Now, Bag Empty, Downstream Occlusion, Door Open, Pump Stopped, KVO, Unit Malfunction |
| Air-in-Line Alarm: |
Air Bubble size of 0.3mL (default value) Can be set from 0.01 to 0.3 mL. |
| Memory: |
User option to retain last Continuous and Piggyback Mode data upon power up. |
| Tubing: |
B.Braun Vista® Administration Sets |
| Modes/Features: |
Continuous and Piggyback Modes, Program Mode for variable infusions, Ramp/Taper Mode for TPN infusions and Delay A Mode for postponing infusion start to finish |
| Operating Temperatures: |
Operating: 64.4º F to 95º F, 10.15 PSI to 15.37 PSI. Non-Operating: -13º F to 131º F, 10.15 PSI to 15.37 PSI |